Simple PC-based calibration using the unit’s push buttons or with the MI70 indicator. Accessed via USB cable connection.
Weather-proof housing (IP66) that ensures continuous, safe data capture and reporting.
Flexible output choices that include Voltage, RS-485, resistive Pt100.
Optional warmed probe technology ensures accurate measurement and better long-term stability in high-condensing environments.
Optional chemical purge function regularly cleans the humidity sensor chip to reduce sensor drift caused by pollution-based chemical vapors — ideal for road and maritime installations.
Optional fast temperature probe for quicker, accurate temperature measurement. Compatible with liquid bath calibration.
Specifications
Technical Specifications
<td”>Operating temperature range<td”>voltage output (0 – 1 V, 0 – 5 V, 0 – 10 V)
resistive Pt100 (4-wire connection)
RS485<td”>3 mA
+0.5 mA
4 mA
max 110 mA
max. 150 mA
RELATIVE HUMIDITY |
|
Measurement range |
0 – 100 %RH |
Accuracy (incl. non-linearity, hysteresis and repeatability) at
+15 to +25 °C (+59 to +77 °F)
-20 to+40 °C (-4 to 104 °F)
-40 to -20 °C (-40 to -4 °F)
+40 to +60 °C (+104 to +140 °F)
-60 to -40 °C (-76 to -40 °F)
Factory calibration
uncertainty (+20 °C /+68 °F) |
±1 %RH (0 – 90 %RH)
±1.7 %RH (90 – 100 %RH)
±(1.0 + 0.008 x reading) %RH
±(1.2 + 0.012 x reading) %RH
±(1.2 + 0.012 x reading) %RH
±(1.4 + 0.032 x reading) %RH
±0.6 %RH (0 to 40 %RH)*
±1.0 %RH (40 … 97 %RH)* |
Response time at +20 °C in still air with a sintered PTFE filter
63%
90% |
20 s
60 s |
TEMPERATURE |
|
Measurement range |
-80 °C to +60 °C (-112 °F to +140 °F) |
Accuracy with voltage output at
-80 to +20 °C
+20 to +60 °C
passive (resistive) output
according to IEC 751 1/3 Class B
RS485 output
-80 to +20 °C
+20 to +60 °C |
±(0.226 – 0.0028 x temperature) °C
±(0.055 + 0.0057 x temperature) °C
±(0.1 + 0.00167 x |temperature|)°C
±(0.176 – 0.0028 x temperature) °C
±(0.07 + 0.0025 x temperature) °C |
Temperature sensor |
Pt100 RTD Class F0.1 IEC 60751 |
Response time with additional temperature probe in 3 m/s air flow
63%
90% |
20 s
35 s |
GENERAL |
|
-80 °C to +60 °C (-112 °F to +140 °F) |
Storage temperature range |
-80 °C to +60 °C (-112 °F to +140 °F) |
Connection |
8-pin male M12 connector |
Connection cables |
3.5, 10, and 30 m |
Cable material |
PUR |
Wire size |
AWG26 |
Service cables |
USB connection cable
MI70 connection cable |
Additional T probe cable length |
2 m |
Housing material |
PC |
Housing classification |
IP66 |
Sensor protection |
sintered PTFE
optional membrane filter |
Weight (probe) |
86 g |
Dimensions |
267 mm L x 40 mm Ø |
Electromagnetic compatibility |
Complies with the EMC standard EN61326-1, Electrical equipment for measurement control and laboratory use – EMC requirement for use in industrial locations |
INPUTS AND OUTPUTS |
|
Operating voltage |
7 to 28 VDC
Note: minimum operating voltage 12 V with 0 to 5 V output and 16 V with 0 to 10 V output, probe heating, chemical purge or XHEAT. |
Outputs |
Average current consumption (+15 VDC, load 100 kOhm)
0 to 1 V output
0 to 10 V output
RS485
during chemical purge
with warmed probe |
Settling time at power-up
voltage output
RS485 |
2 s
3s |
|
|